ENGLISH
INSTRUCTION FOR USE: MEDIVITAL®E
1. FOREWORD
The product complies with the essential requirements of 93/42/EEC Medical Device Directive
and Transportable Pressure Equipment Directive 2010/35 EU. The combination valve is
designed according to EN ISO 10524-3 and EN ISO 10297 standards.
TERM
EN
MEDIVITAL®E
E-GAUGE
DISPLAY
2. INTENDED USE
MediVital
gases. These combination valves together with a gas cylinder form gas packages used either
as gas supply point for medical devices (anaesthetic devices, ventilating devices, incubators
etc.) or for direct gas supply to a patient's breathing mask or cannula.
GCE Combination Valves are intended to be used with the following medical gases:
•
Oxygen
•
Nitrous oxide
•
Air for breathing
•
Helium
•
Carbon dioxide
•
Mixtures of the gases listed above.
MediVital
substances.
2.1. PATIENT PROFILE
Medivital®E supplies medical gas by means of the control flow head in flow range from 0 to 15
(25) l/min.
Saturation of the medical gas must be prescribed by the doctor or an anaesthetic nurse who
has got the medical education, assessed the heath status of the patient and supplies him with
required amount of the medical gas.
Health status - the assesment and decision about the use timing and dosage amount
requirements of the patient is up to the doctor or medical personnel.
Medical personnel - nurse, rescuer, paramedic or carer who has obtained medical education.
2.2. USER PROFILE
Use in the hospital or ambulance:
•
Education: personnel with medical education
Home care use:
Therapy is prescribed by a doctor or personnel who has got a medical education in the field of
breathing support:
•
Knowledge: training of personnel with medical education who has got an ability to read, train
and apply knowledge from the instruction for use.
Training must be performed according to regulations valid in each country.
Important:
Patient must be trained by the doctor or person who has got an education in the life saving.
All manuals and user guides at all-guides.com
MEANING
Combination valve MediVital®E is equipped with an electronic pressure
indicator E-GAUGE
Electronic indicator
Electronical indication device
E Combination Valves are designed to be fitted to gas cylinders used for medical
®
E combination valve is not intended for use with flammable anaesthetics and
®
2/132