Tempering of the device
5.1.2.2. Error alarms - related to the evaluation of the error status of the device. These are
marked with a big warning triangle in the left part of the display (except the "Low battery
voltage" warning)
Low battery charge
If MediVital®E is not used for long periods, check the display of the E-GAUGE every 3 months.
If the symbol
is shown on the display, the valve can still be used, but the valve must
be replaced at the latest within 3 months of detecting this situation. If this is not done, the
operation of the E-GAUGE may be unexpectedly terminated.
Error of the pressure measure
•
The warning is indicated if the E-GAUGE indicates a
rapid change in temperature caused by eg. moving the
device between two rooms with a high difference in
temperatures.
•
Only the pressure parameter is indicated.
•
The warning T symbol is indicated (temperature).
•
A warning is indicated when there is a low battery charge
(2,7V and less).
•
The battery symbol is permanently indicated
•
The battery symbol is static, it does not have an active bar
chart indicator and always includes one bar.
•
The use of the device may continue but at the earliest
cylinder filling, the battery must be changed.
•
The symbol may be indicated even when the battery
charge is OK, but the cylinder was moved to a room with
low ambient temperature. The symbol disappears when
the cylinder warms up again.
•
A warning is indicated when there is a defect of pressure
sensor.
•
The symbol ??? is indicated instead of the pressure or
remaining time.
•
The PS (pressure sensor) is indicate.
•
The bar chart is not indicated.
•
Discontinue the cylinder from service immediately and
send it back to the owner of the filling station. Service
must be performed by a GCE certified service centre.