Replace main actuator
Step 1:
Use a flat screwdriver to remove the star lock
cap and circlips from the upper rod holding the
actuator (from only one side of the pin) and
remove the pin. Do the same operation for the
bottom side of the actuator.
Removing cover
Removing pin
Step 2:
Do the same operation for bottom side of the
actuator.
Removing cover
Removing safety clip
Removing cover
Removing safety clip