seBastian II Service Manual

Device maintaining

All movable joints must be lubricated every six
months or when experiencing squeaks or other
noise. These nodes are:
• axle transport wheels,
• axle actuator mounting,
• jib arm bearing sleeve.
We recommend using any available penetrating
lubricants.
Report all loose connections that are supposed
to be fixed to the manufacturer's authorized
service agent and stop using the product until
the problem is resolved.
This device is equipped with electric actuator
which allows to adjust the height of the jib
arm. In order to ensure proper performance of
the actuators check the actuators for unusual
sounds every 3 months. These unusual sounds
contain rattles, squeaks, creaks. Any perceived
performance deficiency must be reported to
Human Care technical service immediately cea-
sing device operation until remove their causes.
Not less frequently than every 1 years and after
each failure/repair of actuator technical service
must check the lift for electric safety.
The minimal range of the review should include:
• check if there were no mechanical wiring
damages of the device,
• check if there were no mechanical damages
of the controller and the actuators.
See checklist in user manual.
Basic periodic review
One year after production date or every 8000
cycles of using the lift, the indicator lamp light
up.
If this LED is on, operator must check condition
of the lift (cables connections, looseness on the
frame etc.).
Press and hold "up" and "down" button for 5
sec., to turn off the light after review.
6
www.humancaregroup.com
Transport wheels
Actuator mounting
Jib arm bearing sleeve
LED indicator