Step 3:
If everything looks good so far but the device
still doesn't work, for further diagnose, spare
parts will be necessary (hand control, control
box and actuator).
• At first, replace the hand control for new
one (or switch with another device if it's
possible) and check,
• Next, prepare a new control box and connect
it with slots from device,
• Now test what is wrong, control box, hand
control or actuator.
Note – in case of situation that one of actuators
is not working, make sure that problem
is with actuator, by swapping the wires
1 and 2 for actuators in the control box
and use the remote control to test (a
working actuator means a faulty control
box).
Step 4:
To remove control box, just loosen two main
screws and lift up whole holder with control
box.
Loosening two screws
8
www.humancaregroup.com
Removing main holder