Technical specifications (EMC-RF)
Guidance and manufacturer´s declaration – electromagnetic emissions
The HemoCue systems are intended for use in the electromagnetic environment specified below. The customer or the user of the
HemoCue systems should assure that it is used in such an environment.
Emission test
RF emissions
RF emissions
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/flicker emissions
IEC 61000-3-3
Use only cables with the following specification:
USB shielded maximum 2 m
Serial shielded maximum 1.5 m
Compliance
Group 1
Class B
Class A
Complies
49
Electromagnetic environment – guidance
The HemoCue systems uses RF energy only for its
internal function. Therefore, its RF emissions are
very low and not likely to cause any interference in
nearby electronic equipment
The HemoCue systems are suitable for use in all es-
tablishments, including domestic establishments and
those directly connected to the public low-voltage
power supply network that supplies buildings used
for domestic purposes.