Recommended protective distances between portable and mobile
HF telecommunication devices and the TENS eco 2
The TENS eco 2 is intended for operation in an electromagnetic environment in which RF
disturbances are controlled. The customer or the user of the TENS eco 2 can help to avoid
electromagnetic interference by maintaining the minimum distance between portable and mobile RF
telecommunication devices (transmitters) and the TENS eco 2 - depending on the output power of
the communication device, as indicated below.
Rated power of
the Transmitter
W
0,01
0,1
1
10
100
For transmitters whose maximum power rating is not specified in the above table, the recommended
separation distance d in metres (m) can be determined using the equation associated with each
column, where P is the maximum power rating of the transmitter in watts (W) as specified by the
transmitter manufacturer.
NOTE 1
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2
These guidelines may not be applicable in all cases. The propagation of
electromagnetic quantities is affected by absorptions and reflections of buildings,
objects and people
Guidelines and manufacturer's declaration - Electromagnetic emission
The TENS eco 2 unit is intended for operation in an environment as specified below. The customer
or the user of the TENS eco 2 should ensure that it is operated in such an environment.
Störaussendungsmessung Übereinstimmung
HF emissions according to
CISPR 11
HF emissions according to
CISPR 11
Emission of harmonics
according to IEC 61000-3-2
Emission of voltage
fluctuations/ flicker according
to IEC 61000-3-3
Guidelines and manufacturer's declaration - Electromagnetic immunity
The TENS eco 2 is intended for use in the electromagnetic environment specified below. The
customer or the user of the TENS eco 2 should ensure that it is used in such an environment.
Immunity tests
26
Protective distance depending on the transmission frequency
150 kHz to 80 MHz
0,12
0,38
1,2
3,8
12
Group 1
Class B
not applicable
not applicable
IEC 60601 test level
m
80 MHz to 800 MHz
0,12
0,38
1,2
3,8
12
Elektromagnetische Umgebung - Leitfaden
The TENS eco 2 must emit electromagnetic
energy to ensure its intended function.
Neighbouring electronic devices can be
influenced.
The TENS eco 2 is suitable for use in all
facilities including those in residential areas and
those directly connected to a public supply
network that also supplies buildings used for
residential purposes.
Agreement level
800 MHz to 2,5 GHz
0,23
0,73
2,3
7,3
23
Electromagnetic environment
- Guidelines