Download Inhoudsopgave Inhoud Print deze pagina

Settings Menu: Selection Temperature Sensor; Settings Menu: Calibrate Thermostat/External Sensor - Mark OptiTherm+ 06 29 185 Technische Handleiding

Inhoudsopgave

Advertenties

Beschikbare talen

Beschikbare talen

With the  and  button Switch to, to change the mode of operation, and then push the
OK button.
With the  and  button, change the mode of operation in DAY, FROST or ECO and
confirm this by pushing the OK button.
When you have finished the changes and these are confirmed with the OK button, you push
the ESC button to go back to the settings menu.
Now select another item to change or to go back to the user menu by pushing the ESC
button. Push the ESC button again to go back to the main display.

5.8 Settings menu: Selection temperature sensor

With the OptiTherm+ you have the possibility to measure the room temperature by:
1.
Thermostat: sensor built-in to OptiTherm+ itself
2.
External zone: an external sensor connected to each air heater.
3.
External average: an external sensor connected on one or more air heaters. Where the
average value is automatically calculated based on the number of connected sensors.
As standard, the temperature sensor on the OptiTherm+ is used.
When use is made of 1 or more external sensors, you must indicate this in the settings menu
Change selection temperature sensor. See the image below:
back to main screen
2x
ESC
Settings
Date/Time
Overtime setting
Select Temperature sensor
Calibrate
ECO fan function
select
To adjust this:
Select the menu-item Selection temperature sensor and then push the OK button.
Select the sensor by pushing the OK button.
Change the setting and then push the OK button.
You can choose from the thermostat (OptiTherm+ sensor), external zone (each air heater one
own external sensor) or external average (1 or more sensors).
When you have finished with the changes you leave the menu item by pressing the ESC
button and you go back to the settings menu.
Now select another item to change or to go back to the user menu by pushing the ESC
button. Push the ESC button again to go back to the main display.

5.9 Settings menu: Calibrate thermostat/external sensor

When because of unfavourable circumstances there is a deviation in the displayed measured room
temperature and the actual room temperature, you can adjust this deviation by calibrating the
OptiTherm+. Always ensure that the thermostat or external sensor is properly mounted and
connected.
For example:
The measured temperature is 18°C, the screen display on the OptiTherm+ shows 20°C. The
difference in this example is therefore 2°C too high  the correction value must then be set at
-2°C.
ESC
Select Temperature sensor
Sensor
Room Control
OK
select
ESC
Select Temperature sensor
Sensor
Room Control
OK
modify
EN
15

Advertenties

Hoofdstukken

Inhoudsopgave
loading

Gerelateerde Producten voor Mark OptiTherm+ 06 29 185

Inhoudsopgave