GX4 GAS MONITOR
SETUP AND OPERATION CONTINUED
CONNECT THE AIR SUPPLY
FIGURE 4.1
Insert the 3/8" O.D. Air Supply Hose
08-428 into the push-lock fitting on the left
side of the monitor. To disconnect, press the
outside ring of the push-lock and pull the
hose out. Note: Hose can be trimmed to the
desired length.
FIGURE 4.3
Connect the free end of the 3/8" O.D. Air
Supply Hose 08-428 into the one-touch
fitting on the 08-470 regulator.
14
®
FIGURE 4.2
Connect the 08-470 regulator with Quick
Disconnect fitting into one of the outlets
on the Radex or other air source providing
Grade D or better air. Regulate the air from
the air source to between 7 to 80 PSI
(50-550 kPa) with the 08-470 regulator.
Maximum air flow is achieved at 20 PSI. It
is recommended to supply the GX4
15-20 PSI. Max pressure is 80 PSI. Pressure
above 80psi will close the regulator and
there will be no airflow to the monitor.
CONNECTING THE POWER SUPPLY
FIGURE 5.1
Connect the power supply by screwing it
onto the receptacle on the left side of the
unit (next to the air inlet fitting.) Plug the AC
Adaptor into a power source.