GX4 GAS MONITOR
INSTRUCTION MANUAL
Employers: Read this manual, the respirator and the flow control
device Instruction Manual and carry out the Employer Responsibilities
(page 7).
Product users: Read this manual, the respirator and the flow control
device Instruction Manual and follow the Product User Safety
Instructions (page 9).
Manuals are regularly updated. Make sure the most recent version of
manual is available to all users for reference.
Current version of manual and other languages: www.rpbsafety.com/
product/gx4-gas-monitor/
Protecting you for life's best moments
®
ENGLISH
EN
NEDERLANDS
NL
®
2
33

Advertenties

Inhoudsopgave
loading

Samenvatting van Inhoud voor RPB GX4

  • Pagina 1 GX4 GAS MONITOR ® ENGLISH INSTRUCTION MANUAL NEDERLANDS Employers: Read this manual, the respirator and the flow control device Instruction Manual and carry out the Employer Responsibilities (page 7). Product users: Read this manual, the respirator and the flow control device Instruction Manual and follow the Product User Safety Instructions (page 9).
  • Pagina 2: Inhoudsopgave

    GX4 GAS MONITOR ® CONTENTS  EXPLANATION OF SIGNAL WORDS AND SYMBOLS  INTRODUCTION  IMPORTANT SAFETY INFORMATION  PROTECTION PROVIDED AND LIMITATIONS  TYPICAL SUPPLIED AIR SETUP  AIR SOURCE, FITTINGS, AND PRESSURE  EMPLOYER RESPONSIBILITIES  PRODUCT USER SAFETY INSTRUCTIONS ...
  • Pagina 3: Introduction

    RPB Safety, LLC. For technical assistance contact our Customer Service Department at 1-866-494-4599 or email: customerservice@rpbsafety.com...
  • Pagina 4: Gx4 Gas Monitor

    Cartridges are available to detect ® ® Carbon Monoxide (5 or 10ppm depending on the requirements of your region), Oxygen, and Hydrogen Sulfide. Visit www.rpbsafety.com/product/gx4-gas-monitor/ for a list of available gas sensors. The GX4 DOES NOT remove any gasses from the air. ®...
  • Pagina 5: Typical Supplied Air Setup

    ® AIRLINE FILTRATION SYSTEM 3. BREATHING AIR LINE 5. GX4 ® MONITOR FIGURE 1.1 (Continued from page 4) Compressed Gas Association Commodity Specification for Air, G-7.1-1989 Grade D or better air, EN 12021, or AS/NZS 1715 requirements. See Employer Responsibilities section for the other components that need to be sampled.
  • Pagina 6: Air Source, Fittings, And Pressure

    ® AIR SOURCE, FITTINGS, AND PRESSURE AIR SOURCE Locate the air source being supplied to the respirator and being monitored by the GX4 in a ® clean air environment. Make sure the air source is somewhere that vehicles, forklifts, and other machinery are not running near the air inlet, as this will cause carbon monoxide to be drawn into your air supply.
  • Pagina 7: Employer Responsibilities

    ® ® and there will not be enough air flow for the sensors to monitor the air. EMPLOYER RESPONSIBILITIES Your specific responsibilities may vary by location and industry, but in general RPB expects that ® employers will: ...
  • Pagina 8 EMPLOYER RESPONSIBILITIES CONTINUED Train employees in the GX4 ’s use, maintenance, and limitations.  ® Appoint a qualified individual who is knowledgable about the RPB to provide ® ® training: Ex.: ANSI/ASSE Z88.2 Section 8.1 Qualifications of the Respirator Trainer. Anyone providing training shall: a) be knowledgable in the application and use of the monitor(s);...
  • Pagina 9: Product User Safety Instructions

    Airflow stops or slows down.  The alarm in the GX4 or other alarm, such as a strobe and alarm, connected to the GX4 ® ® sounds. Check air source and wait until alarm mode stops. If it continues, verify sensor calibration or change your air source.
  • Pagina 10: Setup And Operation

    ® features and maintain compatibility with the latest cartridges. Check for updates every three months or when installing a new cartridge. Visit http://go.myrpb.com/GX4-Update for information on the latest software and how to upgrade your unit. SENSOR CHECK Checks should be performed when a new cartridge is installed and monthly thereafter.
  • Pagina 11: Wifi Connection

    WIFI CONNECTION To utilize all of the features of the GX4 , it is recommended to connect it to the internet via ® the built-in wireless network transceiver. This connection allows easy firmware upgrades and the unit hosts a website available to devices on that network. Use the website to stream real- time gas levels, search through logs or print reports that can be used for safety records and meetings.
  • Pagina 12: Website

    This ® allows the user to view information about the unit and change settings such as the network connection. This is your portal to access the full potential of the GX4 ® To access the website for your GX4 from your internet connected device: ®...
  • Pagina 13: Unit Setup

    UNIT SETUP MOUNTING TO A RADEX FILTRATION UNIT ® FIGURE 3.1 Position the monitor so that the green brackets on the back line up with the mounting bracket on the side of the RADEX , turn clockwise until the monitor ®...
  • Pagina 14 7 to 80 PSI desired length. (50-550 kPa) with the 08-470 regulator. Maximum air flow is achieved at 20 PSI. It is recommended to supply the GX4 with ® 15-20 PSI. Max pressure is 80 PSI. Pressure above 80psi will close the regulator and there will be no airflow to the monitor.
  • Pagina 15  If connecting to a battery only: connect the black clip to the negative terminal of the battery. Be sure to keep the GX4 as far away from the car, truck, or compressor exhaust as possible. ® Exhaust fumes can interfere with the measurements of the sensors, particularly when the air supply is not yet connected.
  • Pagina 16 GX4 GAS MONITOR ® SETUP AND OPERATION CONTINUED Corrosive Chemicals can Injure Eyes and Skin  Lead-Acid batteries contain sulfuric acid.  Wear complete splash-proof eye protection, face shield and protective clothing such as apron and gloves when working near lead-acid batteries. Always have someone nearby for help.
  • Pagina 17: External Wiring Diagram

    EXTERNAL WIRING DIAGRAM FIGURE 6.3...
  • Pagina 18 FIGURE 7.2 FIGURE 7.1 To turn the unit on, press and hold the “OK” Tip: you can store the AC Adaptor between button until the “rpb” logo appears on the the tabs on top of the mounting bracket. screen. POWER OFF FIGURE 7.3...
  • Pagina 19 Insert the cartridge into any of the 4 slots in the monitor. First locate the top end so Visit rpbsafety.com/product/gx4-gas- the tabs line up, then push the bottom until monitor/ for a list of available sensors.
  • Pagina 20 GX4 GAS MONITOR ® SETUP AND OPERATION CONTINUED REMOVE CARTRIDGE FIGURE 8.5 FIGURE 9.1 Each monitor comes with 3 blank To remove a cartridge, press the cartridges. These should be placed into corresponding release button and the slots that are not currently being used cartridge will “pop up”.
  • Pagina 21: Menus

    The logs are recorded in UTC/GMT Sensors and display according to your settings. Sensor cartridge information on each. Synchronize the Internal Clock Browse each cartridge for: To correct the Time/Date of your GX4 ®  Last date checked  Connect your GX4 to the internet.
  • Pagina 22: Alarms And Settings

    Turn on air flow and adjust the inlet pressure to 15-20psi. a. Check that the flow reading is at least 50. If not, replace the manifold tube. Turn off/disconnect all air flow to the GX4 and record the flow reading shown.
  • Pagina 23 All alarms and silence events are recorded in the event logs. It is recommended that an external alarm system be wired to the Auxiliary outlet. BATTERY The GX4 is designed to be run on external power. It does have a built in Lithium Polymer ®...
  • Pagina 24: Power And Charging

    GX4 GAS MONITOR ® SETUP AND OPERATION CONTINUED To reduce the risk of battery fire or explosion or of fire or electric shock with the charger, follow the below electrical and battery precautions and procedures: Note: FIGURE 10.1 The unit requires 10 seconds to read the battery level after the unit is turned on.
  • Pagina 25: Maintenance

    MAINTENANCE MANIFOLD CHECK AND REPLACE FIGURE 11.2 FIGURE 11.1 In the back of the GX4 , there is a manifold To remove the manifold for inspection, ® (to distribute the air to the sensors) press in the green clip so it can be pulled comprised of a clear tube with end caps.
  • Pagina 26: Replace Mounting Brackets

    GX4 GAS MONITOR ® MAINTENANCE CONTINUED FIGURE 11.5 FIGURE 11.6 After removing the manifold, inspect it from To replace the manifold, insert the tapered all sides to make sure all of the holes are end into the green ring of the push-lock clear and that there is no oil or debris inside fitting inside the back of the monitor.
  • Pagina 27: Cleaning

    Occasionally clean the contacts with cotton swabs dipped in isopropyl alcohol. For black streaks from contacts rubbing, use a pencil eraser or a finger tip. If the cartridge is not registering when put into the GX4 , it is most likely ®...
  • Pagina 28 GX4 GAS MONITOR ® MAINTENANCE CONTINUED Alarms will be disabled. You must confirm the monitor is not in WARNING use. Always store and use in a well ventilated area when working with compressed air and hazardous gases. Avoid contact with eyes, skin and clothing by wearing proper PPE, such as safety glasses and gloves.
  • Pagina 29: Maintenance Schedule

    CARTRIDGE REPLACEMENT The Sensor Cartridges have a two year lifespan from their date of manufacture. After the expiration date their readings may be unreliable. A warning will pop-up on the GX4 when ® the cartridge is going to expire in 30 days and will reset every day until it expires. Once the cartridge expires a message will pop every 6 hours and the AUX will not output the OK signal.
  • Pagina 30: Parts And Accessories

    GX4 GAS MONITOR ® PARTS AND ACCESSORIES FIGURE 14.1...
  • Pagina 31: Parts List

    Use only exact, authentic RPB replacement parts (marked with ® the RPB logo and part number), and only in the specified configuration. Using incomplete ® or inappropriate equipment, including the use of counterfeit or non-RPB parts, can result ® in inadequate protection.
  • Pagina 32: Warranty And Liability Statement

    Safety is not liable for damage, loss or expense resulting from the failure to give advice ® or information or the giving of incorrect advice or information, whether or not due to RPB ® Safety’s negligence or that of its employees, agents or subcontractors.
  • Pagina 33: Nederlands

    Unit 2, Teardrop Industrial Park, London Road, Swanley, BR8 8TS, United Kingdom T: +44-800-689-5031 E: sales@rpbsafety.co.uk rpbsafety.com Copyright ©2019 RPB Safety, LLC. Alle rechten voorbehouden. Alle inhoud op deze website is beschermd door de Amerikaanse auteursrechtwetgeving en mag niet worden...
  • Pagina 34: Belangrijke Veiligheidsinformatie

    GX4 GAS MONITOR ® gereproduceerd, gedistribueerd, overgedragen, weergegeven, gepubliceerd of uitgezonden zonder de voorafgaande schriftelijke toestemming van RPB Safety, LLC. U mag geen handelsmerk, auteursrecht of andere kennisgeving van kopieën van de inhoud wijzigen of verwijderen. Alle handelsmerken, servicemerken en logo’s die in deze publicatie worden gebruikt, zowel geregistreerd als niet-geregistreerd, zijn de handelsmerken, servicemerken of logo’s van hun respectieve eigenaars.
  • Pagina 35: Bescherming En Beperkingen

    CGA D-Klasse specificaties voor ® ® luchtkwaliteit zoals gespecificeerd door OSHA, of deze overtreft. Raadpleeg de volgende normen en voorschriften, afhankelijk van de locatie waar de GX4 wordt gebruikt, voor meer ® informatie: ...
  • Pagina 36: Luchtbron, Fittingen En Druk

    GX4 GAS MONITOR ® LUCHTBRON, FITTINGEN EN DRUK LUCHTBRON Positioneer de luchtbron van het ademhalingstoestel die door de GX4 wordt gecontroleerd ® in een omgeving met schone lucht. Zorg ervoor dat de luchtbron ergens is waar voertuigen, vorkheftrucks en andere machines niet in de buurt van de luchtinlaat aan staan, omdat hierdoor koolmonoxide in uw ademlucht wordt aangetrokken.
  • Pagina 37: Verantwoordelijkheden Van De Werkgever

    Regelaar 08-470 kan worden gebruikt om de luchtdruk die de GX4 ® ® binnengaat te reguleren. Als er meer dan 80 PSI wordt geleverd aan de GX4 , klinkt het ® alarm voor een lage flow en is er onvoldoende luchtstroom voor de sensoren om de lucht te controleren.
  • Pagina 38 OSHA 29 CFR 1910.134, EN 12021, AS/NZS 1715 of lokale normen zoals vereist.  Train medewerkers in het gebruik, onderhoud en de beperkingen van de GX4 ® Benoem een gekwalificeerd persoon met voldoende kennis over de RPB ®...
  • Pagina 39 ONDERHOUD  Plaats de GX4 nooit op hete oppervlakken. Breng geen verf, oplosmiddelen, kleefstoffen ® of zelfklevende etiketten aan, tenzij aan de hand van instructies van RPB . Bepaalde ® chemicaliën kunnen schadelijk zijn voor dit product.  Reinig met een mild reinigingsmiddel en een zacht doekje, en zorg ervoor dat de monitor niet nat wordt.
  • Pagina 40: Montage En Gebruik

    Controleer elke drie maanden op updates of wanneer u een nieuw patroon installeert. Ga naar http:// go.myrpb.com/GX4-Update voor informatie over de nieuwste software en over het upgraden van uw eenheid.
  • Pagina 41: Hotspot-Modus

    Hier kan ® de gebruiker informatie over het apparaat bekijken en instellingen zoals de netwerkverbinding wijzigen. Dit is uw portal voor toegang tot het volle potentieel van de GX4 ® Om toegang te krijgen tot de website voor uw GX4 vanaf uw apparaat (dat op het internet is ®...
  • Pagina 42: Gx4 Gas Monitor

    Gebruik de muurbeugel (P/N: 08-231) en de Maximale luchttoevoer wordt bereikt bij juiste schroeven om de bevestigingsbeugel 20 PSI. Het wordt aanbevolen om de GX4 ® aan een geschikte muur te bevestigen. De te voorzien van 15-20 PSI. De maximale verzonken gaten zijn 7/32 “(5 mm) wijd en...
  • Pagina 43 Sluit de GX4 niet rechtstreeks aan op 18V- of 24V-batterijen. Om ® verbinding te maken met een batterij die hoger is dan 12V, raadt RPB ten zeerste aan een ® geschikte DC-DC-omvormer te gebruiken. Een omvormer zal ook helpen beschermen tegen stroomstoten en wordt ook aanbevolen voor gebruik met 12V-batterijen.
  • Pagina 44 SLUIT HET EXTERNE ALARM AAN AFBEELDING 6.1 OP PAGINA 16 Als u een extern alarmapparaat gebruikt, steek de stekker in de hulpvoeding (AUX). Het gebruik van een extern alarm wordt aanbevolen als de GX4 niet door de gebruiker kan ®...
  • Pagina 45: Extern Aansluitschema

    750 mA Niet OK Sluit de relaisspoelen (12 VDC, SPDT) aan op poort 1 en 5 van de GX4-hulpvoeding. Sluit uw apparatuur aan op de "OK" of "Niet OK" -posities van het circuit om gebruikers te waarschuwen voor mogelijke gevaren.
  • Pagina 46 Configureer een WiFi-verbinding of wijzig scherm. Na het aftellen gaat het alarm af geavanceerde IP-instellingen. als de waarschuwing niet is gewist voordat Wanneer u de GX4 op een netwerk ® het aftellen eindigt. Druk op de knop aansluit, wordt u mogelijk gevraagd onder het woord “REMOVE”...
  • Pagina 47: Aanpassing Lage Toevoer Instelpunt

    De logboeken worden vastgelegd in Alarm en Hulp UTC/GMT en worden weergegeven volgens uw instellingen. Test het alarm en controleer de status van de hulpapparatuur. Synchroniseer de Interne Klok Om de Tijd/Datum van uw GX4 aan te Apparaatsoftware ® passen: Fabrieksreset van firmware-instellingen. ...
  • Pagina 48 Zet de luchttoevoer open en pas de inlaatdruk aan tot 15-20 PSI. a. Controleer of de toevoerwaarde ten minste 50 is. Zo niet, vervang dan het spruitstuk. Schakel alle luchttoevoer naar de GX4 uit/koppel deze los en noteer de ®...
  • Pagina 49 Lithium-Polymeerbatterij die bedoeld is voor back-updoeleinden in het geval van een stroomstoring. De monitor kan tot 10-12 uur op de batterij werken. Het wordt niet aanbevolen om de GX4 ® regelmatig of gedurende lange tijd op batterij-stroom te laten werken.
  • Pagina 50: Onderhoud

    VERVANGING Aan de basis van het spruitstuk zit een AFBEELDING 11.1 OP PAGINA 25 push-lock fitting met een groene ring. Trek Aan de achterkant van de GX4 bevindt ® het spruitstuk los terwijl u op de ring drukt. zich een spruitstuk (om de lucht naar de AFBEELDING 11.5 OP PAGINA 26...
  • Pagina 51 Gebruik een potloodgum of een vingertop om zwarte strepen van contactbewegingen te verwijderen. Als het patroon niet registreert wanneer deze in de GX4 wordt geplaatst, is het zeer ®...
  • Pagina 52  Controleer Luchtfles(sen) die specifiek zijn voor het gas van de patronen die moeten worden gecontroleerd.  Kalibratiekit (08-451) om de flessen aan te sluiten op de GX4 ® Alarmen worden uitgeschakeld. U moet bevestigen WAARSCHUWING dat de monitor niet in gebruik is. Altijd opbergen en gebruiken in een goed geventileerde ruimte bij het werken met ademlucht en gevaarlijke gassen.
  • Pagina 53 De sensorpatronen hebben een levensduur van twee jaar vanaf de fabricagedatum. Na de vervaldatum kunnen hun waarden onbetrouwbaar zijn. Er verschijnt een waarschuwing op de GX4 wanneer het patroon over 30 dagen verloopt en dit wordt elke dag opnieuw vermeld ®...
  • Pagina 54 Gebruik alleen exacte, authentieke RPB ® vervangingsonderdelen (gemarkeerd met het RPB -logo en onderdeelnummer) en ® alleen in de aangegeven opstelling. Gebruik van onvolledige of ongeschikte uitrusting, waaronder het gebruik van nagemaakte of niet-RPB -onderdelen, kan resulteren in ® onvoldoende bescherming.
  • Pagina 55: Beperkte Garantie

    ® ® tevredenheid te bepalen of een geclaimd defect onder deze beperkte garantie valt. Als er defect optreedt dat onder de garantie valt, zal RPB naar eigen goeddunken het defecte ® Product (of een onderdeel van het product) repareren of vervangen. Deze ‘reparatie of vervanging’...
  • Pagina 56: Other Products

    Look no further than the . From the searing heat ® ™ of an Arizona summer to a severe Scandinavian winter the RPB ® ™ will keep you comfortable. NOVA 3 RESPIRATOR ®...

Inhoudsopgave