Download Inhoudsopgave Inhoud Print deze pagina

Vega VEGAPULS C 22 Handleiding pagina 51

Verberg thumbnails Zie ook voor VEGAPULS C 22:
Inhoudsopgave

Advertenties

Read Power Operation Mode
Command
Response
aXRPOM!
a<value1><value2> <CR><lf>
Example:
Command
Response
0XRPOM!
0+0<CR><LF>
Write Power Operation Mode
Command
Response
aXWPOM <va-
a<value1><value2> <CR><LF>
lue1>!
Read distance unit
Command
Response
aXRDU!
a<value1><CR><LF>
Write distance unit
Command
Response
aXWDU<value1>! a<value1><value2> <CR><LF>
Example:
Command
Response
0XWDU+0!
0+0+000<CR><LF>
0XWDU+4!
0+0+136<CR><LF>
VEGAPULS C 22 • SDI-12
Description
a: sensor address
<value1>: power operation mode, +0 = low power mode,
+1 = normal power mode
<CR><LF>
Description
Description
a: sensor address
<value1>: power operation mode, +0 = low power mode,
+1 = normal power model
<value2>: VVO-Status eee
<CR><LF>
Description
a: sensor address
<value1>: distance unit +0 = unit in [m], +1 = unit in [ft],
+2 = unit in [mm], +3 = unit in [in]
<CR><LF>
Description
a: sensor address
<value1>: distance unit +0 = unit in [m], +1 = unit in [ft],
+2 = unit in [mm], +3 = unit in [in]
<value2>: VVO-Status +eee
<CR><LF>
Description
Valid data
No valid data
Current value is returned with a status 136