the BENNING MM 7.
-
Contact the measuring points with the safety test leads. If the measuring
lead resistance between the COM socket  and the socket  for V, Ω, Hz,
ºC, ºF,
on the BENNING MM 7 is lower than 30 Ω the built-in buzzer is
activated.
See fig. 8:
Continuity testing with buzzer
8.6 Capacitance measurement
Discharge capacitors fully before measurement! Never apply
voltage to the sockets for capacitance measurement as this
may cause irreparable damage to the unit. A damaged unit may
represent an electrical hazard!
-
With the rotating switch , select the desired function on
BENNING MM 7 Determine the polarity of the capacitor and discharge it
completely.
-
Connect the black safety test lead to the COM socket  on the
BENNING MM 7.
-
Connect the red safety test lead to the socket  for V, Ω, Hz, ºC, ºF,
the BENNING MM 7.
-
Contact the discharged capacitor with the safety test leads observing cor-
rect polarity. Read the measurement value on the digital display  of the
BENNING MM 7.
See fig. 9:
Capacity measurement
8.7 Frequency measurement
-
With the rotating switch , select the desired function (Hz) on the
BENNING MM 7.
-
Connect the black safety test lead to the COM socket  on the
BENNING MM 7.
-
Connect the red safety test lead to the socket  for V, Ω, Hz, ºC, ºF,
on the BENNING MM 7. Remember the minimum sensitivity for frequency
measurements using the BENNING MM 7!
-
Contact the measuring points with the safety test leads and read the meas-
urement result on the digital display  on the BENNING MM 7.
See fig. 10:
Frequency measurement
8.8 Temperature measurement
-
With the rotating switch , select the desired function (°C or °F) on the
BENNING MM 7
-
Connect the adapter for the temperature sensor correctly to the socket
COM  and V, Ω, Hz, °C, °F,
-
Connect the temperature sensor (type K) to the adapter.
-
Place the contact point (end of the sensor lead) on the point to be
measured. Read the measurement value on the digital display  of the
BENNING MM 7.
See fig. 11:
Temperature measurement
9.
Maintenance
Before opening the BENNING MM 7, ensure that it is not
connected to a source of voltage! Electrical hazard!
Any work required on the BENNING MM 7 when it is under voltage must be done
only by a qualified electrician. Special steps must be taken to prevent accidents.
Before opening the BENNING MM 7, remove it from all sources of voltage as follows:
-
First remove both safety test leads from the measurement points.
-
Remove both safety test leads from the BENNING MM 7
-
Turn the rotating switch  to "OFF".
9.1 Securing the unit
Under certain circumstances, the safety of the BENNING MM 7 can no longer
be guaranteed. This may be the case if:
-
there are visible signs of damage on the unit,
-
errors occur in measurements,
-
the unit has been stored for a long period of time under the wrong condi-
tions, and
-
if the unit has been subjected to rough handling during transport.
In these cases, the BENNING MM 7 must be switched off immediately, removed
from the measuring points and secured to prevent it from being used again.
9.2 Cleaning