With S-Guard
Before putting on the loupes, select your inter-pupillary distance (PD) or an average
value on the scale. Unscrew the retaining screw and slide the eyepieces to one side
symmetrically.
Unscrew the fixing screw (4) and locate the nosepad on the bridge of the nose.
Adjust the eyepieces so that they nearly touch the protective lenses.
Shine the ML4 LED onto an object at your normal working distance and adjust the
eyepieces so that the object is centred in both eyepieces. After adjustment, re-tighten
the screw (4).
The loupes are correctly adjusted when you see a clearly-defined, round, shadow-free
field of view. The slider (5) allows up or down adjustment of size of the light field.
LoupeLight 2 and MicroLight 2
Use the SC1 connecting cable to connect the
LED light to the mPack mini (note the shape of
the plug!).
Use the knob of the mPack mini to switch the
LED light on and adjust the head of the light
vertically such that the examination field is
illuminated centrally.
Turing the swivel lever at the mPack mini allows
continuous adjustment of the brightness.
Fasten the cable on your clothes using
the support clip provided (see Fig. 11).
LoupeLight 2
Adjust the binocular loupe for the intended application. Be sure the eyepieces are
positioned as symmetrically as possible on the optics holder (take off the loupe and
check this), as otherwise the spot of light will appear shifted to the side.
1
MicroLight 2
In order that the MicroLight 2 remains in place in every gradually and freely selectable
position, it may be necessary to retighten the screws using a screwdriver (see Fig. 13).
Fig. 13 Tighten the screws
12/76
LoupeLight 2 on loupe mount with i-View
If required, a steri-lever (12) can be attached
to one end (1) of the loupe mount so that the
binocular loupe can be swivelled indepen-
dently of the light (only HR/HRP Loupes).
Fig. 12 Using the steri-lever
MED 113192
2017-01-19
Fig. 11 Cable fastening with
support clip
.
®