You can check a total of 40 temperature readings. When the maximum number of records
is exceeded, the oldest recorded data will be overwritten.
The icon on the display indicates the location from which the stored temperature measurement
was taken. The memory function does not include ambient temperature measurement.
SOUND AND LIGHT INDICATOR
Temperature range
Forehead temperature
32.00C - 37.50C /
89.60F - 99.50F
37.60C - 38.00C /
99.60F - 100.40F
38.10C - 42.90C /
100.50F - 109.20F
Ear temperature
32.00C - 37.50C /
89.60F - 99.50F
37.60C - 38.00C /
99.60F - 100.40F
38.10C - 42.90C /
100.50F - 109.20F
Ambient temperature
00C - 400C /
32.00F - 104.00F
When the measured forehead or ear temperature is within the range 32.00C to 37.50C
(89.60F to 99.50F), the thermometer will emit a long sound signal and the indicator will be
illuminated in green, indicating that the temperature is normal.
When the measured forehead or ear temperature is within the range of 37.60C to 42.90C
(99.60F to 109.20F), the thermometer will give 6 short sound signals (2 short sound
signals each time) and the indicator will be illuminated in orange or red, indicating that
the temperature is too high.
TROUBLESHOOTING
Problem
The device does not
switch on.
Sound (does not apply to silent mode)
One long sound signal
6 short sound signals (2 short sound
signals each)
6 short sound signals (2 short sound
signals each)
One long sound signal
6 short sound signals (2 short sound
signals each)
6 short sound signals (2 short sound
signals each)
One long sound signal
Possible cause
1. Battery level is too low.
2. Batteries have been placed
in the device with the wrong
polarity.
3. The device has been
damaged.
Light indicator
Green
Orange
Red
Green
Orange
Red
-
Solution
1. Replace the batteries.
2. Ensure that the batteries
have been correctly inserted
in the device.
3. Contact an authorised
service centre.