- Fill a container of suitable size to hold all of the parts to be disinfected with a
solution of potable water and disinfectant, according to the proportions indicated
on the packaging of the disinfectant.
- Completely immerse each part in the solution, taking care to avoid the formation of
air bubbles on the parts. Leave the parts immersed for the amount of time indicated
on the packaging of the disinfectant associated with the concentration chosen for
the solution.
- Remove the disinfected parts and rinse with plenty of warm potable water.
- Dispose of the solution following the instructions provided by the disinfectant
manufacturer.
If you want to also perform the STERILISATION, jump to the STERILISATION paragraph
After having disinfected the accessories, shake them vigorously and place them on
a paper towel. Alternatively, dry them with a jet of hot air (for example, from a hair
dryer).
STERILISATION
The accessories that can be sterilised are 1A-1B-1C-2-3-4-5.
Equipment: Fractionated vacuum overpressure steam steriliser in accordance with EN
13060.
Directions: Wrap every single part to be treated with a sterile barrier system or
packaging in accordance with Norm EN 11607. Place the packed components in the
steam steriliser. Run the sterilisation cycle according to the operating instructions of
the device by selecting a temperature of 121°C and a time of 10 minutes first.
Storage: Store the sterilised parts as per the instructions for use of either the sterile
barrier system or packaging.
The sterilisation procedure is validated in its conformity to ISO 17665-1.

NASAL WASH SYSTEM TECHNICAL FEATURES

Mod.: Rhino Clear® Ergo Plus
Operating Conditions:
Temperature: min 10°C; max 40°C
Relative humidity: min 10%; max 95%
Atmospheric pressure: min 69KPa; max 106KPa
Capacity: medication maximum 10 ml
Type BF applied parts are: patient accessories (1A-1B-1C)
Compliant with Directive 93/42 EEC-
Family of FLAEM compressors
F400
F700/F1000
F2000
Tests performed with 0.9% NaCl saline solution, according to the Flaem internal procedure
(1)
Storage conditions:
Temperature: min -25°C; max 70°C
Relative humidity: min 10%; max 95%
Atmospheric pressure: min 69KPa; max 106 KPa
10 ml Delivery Time
2 minutes (approx)
2 minutes (approx)
2 minutes (approx)