Troubleshooting

Symptom
Unit does not work
"Low pressure" alarm
Mattress does not
pump up
Patient bottoms out
No alternating pressu-
re takes place

Technical data

UNIT
Power supply
230 V / 50 hz;
protection class 1; IPX0;
type BF, AP/APG NO
Power consumption max. 25 Watt
Weight
3,9 kg
Dimensions
36.9 x 13.5 x 25 cm
Pressure range
10 - 50 mbar
Alternating cycle
10/15/20/25 min.
Drive Medical GmbH & Co. KG | As per: 28.07.2023 | Errors and alterations excepted
Remedial action
• Check whether the plug is correctly inserted. Switch on the unit again.
• If the LED on the On/Off switch does not light up, check the voltage of the outlet or
insert the plug into another outlet.
• If the LED on the On/Off switch lights up but the unit does not work, please contact
your specialist dealer.
• Check whether the connection hose on the unit is correctly inserted.
• The CPR valve must be closed.
• Ensure that all cells are correctly connected to each other and that the connection
hoses are not twisted or kinked.
• This indicates excessively low pressure in the air chambers. Increase the user weight
on the unit and perform a hand check to check whether the patient is sinking down.
• Check air filter (located above the on/off switch on the side of the unit); please clean
or replace it.
• If the unit continues to generate too little pressure, the unit must be replaced. Contact
your specialist dealer
• Ensure that the air supply hoses are not kinked or twisted.
• The static function must be switched off. When the static function is switched on, no
alternating pressure takes place.
• Ensure that the autofirm and seat functions are switched off.
MATTRESS REPLACEMENT SYSTEM
Dimensions 200 x 88 x 20 cm
cells
20 cells of which 9 are air flow
cells, 3 are static head cells
Cell height 20 cm
Cover
Nylon/PU, breathable, washable
cells Nylon/PU;
ENVIRONMENTAL PARAMETERS
Temperatures
Operation
10 - 35°C
Storage
-15 – 50°C
Shipment
-15 – 70°C
Relative humidity
Operation
20 – 80% => non-condensing
Storage
10 – 90% => non-condensing
Instructions for use FLEX WAVE
EN
19