Air fl ow
The alternating pressure system has nine air fl ow cells.
This group of air chambers is microperforated at the
edge and in this way, it ensures that the alternating pres-
sure system is actively supplied with air.

CPR (Cardiopulmonary resuscitation)

Before resuscitation can/may be performed on the sys-
tem, the system must be fully defl ated.
For rapid defl ation, pull down on the red tab printed with
„CPR" at the head end. The tab is on the same side as
the hose system. The valves open and air escapes from
all of the cells.
To supply the system with air once again, close the CPR
valve again by pressing it fi rmly and securely fasten the
tab marked „CPR" once again (photo 8).
8

Transport function

Remove the hose system from the unit and promptly
close it with the attached cap. Now the mattress is in
transport mode. To return to normal operation, remove
the cap on the connection hose and connect it back to
the unit and start the unit.
Note: When the Med Aire FlexWave is in transport
mode, it is possible to leave patients on the system for
up to 12 hours. However, at this time, we recommend
regularly monitoring the patient and the mattress repla-
cement system.

Power failure

In the event of a power failure, set the system to trans-
port mode. See section on transport function.
Drive Medical GmbH & Co. KG | As per: 28.07.2023 | Errors and alterations excepted

Alarm function

Low pressure
The „Low Pressure" LED is lit whenever the unit detects
too little air in the cells. An acoustic signal sounds at the
same time if it has not been switched off beforehand by
pressing the „Mute" button.
Once the set pressure has been reached in the cells
again, the display goes out and the acoustic signal stops.
The „Normal Pressure" LED is lit.
Important: During the fi lling phase, the „Low Pressure"
LED is lit; once the pressure in the air chambers has
reached the set user weight, the „Low Pressure" LED
goes out.
System failure/service display
If the unit detects an internal error, the „System Failure"
LED is continuously lit and an acoustic signal additionally
sounds. This can be muted using the „Mute" button. The
„System Failure" LED will still be lit. Please notify your
specialist dealer.
Power failure
If there is a power failure or if the unit is disconnected
from the power supply, the „Power Failure" LED begins
to blink and an acoustic signal sounds.
This alarm stays on until the unit is reconnected to the
mains or the On/Off switch is operated.

Cleaning, disinfection and care

Mattress overlay system
The mattress overlay system should not be saturated or
soaked*, however it can be cleaned by hand. To do this,
use a moist (not wet) cloth and a mild standard cleaning
and disinfectant solution.
* There is a risk of liquid getting into the cell chambers
and hoses.
Unit
• Before cleaning, disconnect the unit from the power
supply. During cleaning, please ensure that no fl uids/
cleaning solutions come into contact with the plug.
• The unit should not be saturated or soaked.
• The unit may be cleaned by hand. To do this, use a
moist (not wet) cloth and a mild household cleaning
and disinfectant solution.
• Do not use any solutions which contain phenol or abra-
sive cleaners. This can corrode the surface of the unit.
• Allow the unit to dry thoroughly.
• Perform a function test after the device is thoroughly
dry.
EN