Cal Checker
EN
not TB350 
1) 
4.6 Verification
4.6.1 Verification Overview
Verification
1
2
Verification Standard:
T-CAL 5.0 NTU
3
Pass/Fail Criteria:
4
Standard Set:
7
5
6
Description
1 Application Bar
2 Verification Standard
3 Pass / Fail Criteria
4 Standard Set
5 Settings Button
26
4 Operation
On
(Default)
Off
12:36
± 10 %
None
Function
Breadcrumb to help the user through the process steps.
Value of Standard to be measured.
Tolerance used for determining if a verification passes or
fails.
Unique identifier of standard set.
Used to change settings.
During the calibration procedure, if an
individual standard is out of tolerance, the
user must correct the standard to proceed
or cancel the calibration procedure.
Feature prevents user from a calibration
error due to an incorrect standard. The log
screen will have a line item for Calibration
Checker and it will show an "On" or
"Enabled" state.
During the calibration procedure, if an
individual standard is out of tolerance,
user will not be notified that a calibrator
is out of Tintometers predefined range.
The log screen will have a line item for
Calibration Checker and it will show an
"Off" or "Disabled" state.
EN   TB 350