2. Prepare Vial:
Oil sample vial if necessary.
3. Insert Standard:
Insert standard value specified on screen.
User will insert series of various turbidity standards to complete
calibration.
4. Modal:
Represents active process.
5. Log Summary:
Press Save Button to save log.
Press Cancel Button to return to beginning of calibration
process.
4.5.3 Calibration Settings
Press the Settings Button to access the settings mode.
Icon
Description
Cal Curve
Standard Set
Reminder
1)
Default Cal
Post Verify
EN   TB 350
Option
Function
T-CAL
Selects prepared stabilized formazine
®
standard at pre-defined concentrations
User
Selects user prepared formazine standard
Formazin
at pre-defined concentrations
Default
Used if no unique identifier is set.
Unique
Unique identifier of standard set.
On
Reminds user that calibration is due.
Option to force Calibration before
additional measurements occur.
Off
Default setting.
 
Option to return to factory calibration
setting.
On
Requires user to perform a Verification
after calibration is complete.
Off
Default setting.
4 Operation
EN
25