4.5 Calibration
4.5.1 Calibration Overview
Calibration
1
EN
2
Calibration Curve:
T-CAL 0-4000 NTU
3
Standard Set:
None
6
4
5
Description
1 Application Bar
2 Calibration Curve
3 Standard Set
4 Settings Button
5 Home Button
6 Next Button
4.5.2 Performing a Calibration
Please prepare the T-CAL
calibration. Make sure that the T-CAL
blemishes.
• T-CAL
standards for TB 350 IR: 20 / 800 / 4000 NTU
®
• T-CAL
standards for TB 350 WL: 20 / 800 / 2000 / 4000 NTU
®
24
4 Operation
12:36
Function
Breadcrumb to help the user through the process steps.
Specifies calibration type.
Unique identifier of standard set.
Used to change settings.
Returns user to home screen.
Progress to next screen.
standards included in the scope of delivery before
®
1. Zero Adjustment:
Sets zero standard.
vials are free of stains, scratches or
®
EN TB 350