N.
Read/
Data
Write
Type
204
R
N
Pump
temperature
in °C
205
R
N
Pump status
206
R
N
Error code
211
R
N
Controller
Heatsink
Temperature
(°C)
216
R
N
Controller Air
Temperature
(°C).
224
R
A
Pressure
reading
Format = X.X
E XX
226
R
N
Rotation
Frequency
(rpm)
300
R
N
Cycle time in
minutes
(zeroed by
the reset
command)
301
R
N
Cycle
number
(zeroed by
the reset
command)
302
R
N
Pump life in
hours
(zeroed by
the reset
command)
320
Reserved to Varian service
to
399
400
R
A
CRC
EPROM
(QE)
402
R
A
CRC Param.
(PA)
404
R
A
CRC
Parameter
structure
500
Reserved to Varian service
503
R/W
N
RS 485
address
504
R/W
L
Serial type
select
Description
Admitted
Values
0 to 70
Stop = 0
Waiting intlk = 1
Starting = 2
Auto-tuning = 3
Braking = 4
Normal = 5
Fail = 6
Bit description:
see the following
figure
0 to 999999
0 to 9999
0 to 999999
QE8XXXX
(where "XXXX"
are variable)
PA8XXXX
(where "XXXX"
are variable)
"XXXX"
0 to 31
(default = 0)
0 = RS 232
1 = RS 485
(default = 0)
1. Automatic means that when the controller stops, the vent
valve is opened with a delay defined by window n. 126;
when the controller starts, the vent valve is immediately
closed.
2. On command means that the vent valve is opened or
closed by means of window n. 122.
7
6
TOO HIGH LOAD
SHORT CIRCUIT
OVERVOLTAGE
AUX FAIL
Window N. 206 Bit Description
85
TECHNICAL INFORMATION
NOTES
5
4
3
2
1
0
NO CONNECTION
PUMP OVERTEMP.
CONTROLL. OVERTEMP.
POWER FAIL
87-900-986-01 (C)