CHARACTER
ASCII
"A"
41h
CRC
-
"B"
42h
CRC
-
"C"
43h
CRC
-
"D"
44h
CRC
-
"E"
45h
CRC
-
"F"
46h
CRC
-
"G"
47h
CRC
-
"H"
48h
"I "
49h
CRC
-
"J"
4Ah
CRC
-
"K"
4Bh
CRC
-
"ACK"
06h
CRC
-
"NACK"
15h
CRC
-
Answer = after a request from the host, the
Turbo-V 81-AG controller will answer in one of the following
ways:
•
ACK
•
NACK
•
Message
When the Request is "A, "B", "C", "D", "F", the Turbo-V 81-
AG controller will Answer the ACK or NACK.
CONVERTED ASCII
FOR CRC CALCULATION
-
BFh
-
BEh
-
BDh
-
BCh
-
BBh
-
BAh
-
B9h
-
-
B7h
-
B6h
-
B5h
-
FAh
-
EBh
When the Request is "E, the Answer will contain the
complete set of the following parameters:
BYTES
1
MSB
LSB
- - - - X X X X
2÷5
6÷9
10÷11
12
13
14÷17
18÷19
20
21
22
*The values for current and voltage are given as numbers,
scaled from 0 to 255, where 0 corresponds to 0 V and 255
to the full scale voltage (130 V) or current (2.5 A).
When the Request is "G", the Answer will contain a string
of 11 characters with the following parameters:
BYTES
1-2
Pump cycles number
(integer coded in 2 bytes)
3
Speed threshold
4-7
Run up time in seconds
(long coded in 4 bytes)
8
Deat time
(0 = NO 1 = YES)
9
Reserved
10
Soft Start mode
(0 = NO 1=YES)
11
CRC
When the configuration parameters have to be changed,
send a Request string a string with 9 characters of the
following type:
"H" + .... data .... + CRC
80
TECHNICAL INFORMATION
MEANING
0
= STOP
1
= WAITING INTERLOCK
2
= STARTING
3
= NORMAL OPERATION
4, 5 = HIGH LOAD
6
= FAILURE
7 = APPROACHING LOW SPEED
Cycle Time
Pump Life
Pump Temperature
Current*
Voltage*
Frequency
Cycle #
R1 Status
R2 Status
CRC
MEANING