Window Protocol - Agilent Technologies Turbo-V 81-AG Bedrijfshandleiding

Inhoudsopgave

Advertenties

The following parameters can be changed:
BYTES
1
Speed threshold
2-5
Run up time in seconds
(long coded in 4 bytes)
6
Deat time
(0 = NO 1 = YES)
7
Reserved
8
Soft Start mode
(0 = YES 1 = NO)
9
CRC
When the Request is "I" (Status readings), the Answer
will contain a string of 2 characters with the following
parameters:
BYTES
1
MSB
LSB
- - X X X X X X
0
= STOP
1
= WAITING INTERLOCK
2
= STARTING
3
= NORMAL OPERATION
4, 5 = HIGH LOAD
6
= FAILURE
7 = APPROACHING LOW SPEED
R2 status 0 = OFF
R1 status 0 = OFF
2
CRC
When the Request is "J" (Numerical readings), the
Answer will contain a string of 5 characters with the
following parameters:
BYTES
1
Current (0-255 scaled)
2
Voltage (0-255 scaled)
3
Rotational speed KRPM
4
Pump temperature °C
(0-254 temperature reading, 255 = fail)
5
CRC
When the Request is "K" (Counters readings), the
Answer will contain a string of 11 characters with the
following parameters:
BYTES
1-4
Cycle time
5-8
Pump life
9-10
Cycle number
11
CRC
On request a sample program in QBasic language is
available by Varian.
MEANING
MEANING
1 = ON
1 = ON
MEANING
MEANING

Window Protocol

Description
Communication Format
8 data bit
no parity
1 stop bit
baud rate: 600/1200/2400/4800/9600 programmable
Communication Protocol
The communication protocol is a MASTER/SLAVE type
where:
Host = MASTER
Controller = SLAVE
The communication is performed in the following way:
1. the host (MASTER) send a MESSAGE + CRC to the
controller (SLAVE);
2. the controller answer with an ANSWER + CRC to the
host.
The MESSAGE is a string with the following format:
<STX>+<ADDR>+<WIN>+<COM>+<DATA>+<ETX>+<CRC>
where:
When a data is indicated between two quotes ('...') it
means that the indicated data is the corresponding
ASCII character.
− <STX> (Start of transmission) = 0x02
− <ADDR> (Unit address) = 0x80 (for RS 232)
<ADDR> (Unit address) = 0x80 + device number (0
to 31) (for RS 485)
− <WIN> (Window) = a string of 3 numeric character
indicating the window number (from '000' to '999');
for the meaning of each window see the relevant
paragraph.
− <COM> (Command) = 0x30 to read the window,
0x31 to write into the window
<DATA> = an alphanumeric ASCII string with the
data to be written into the window. In case of a
reading command this field is not present. The field
length is variable according to the data type as per
the following table:
DATA TYPE
Logic (L)
Numeric (N)
Alphanumeric
(A)
− <ETX> (End of transmission) = 0x03
− <CRC> = XOR of all characters subsequent to
<STX> and including the <ETX> terminator. The
value is hexadecimal coded and indicated by two
ASCII character.
81
TECHNICAL INFORMATION
NOTE
FIELD
VALID
LENGTH
CHARACTERS
1
'0' = OFF
'1' = ON
6
'-', '.', '0' . . . '9' right
justified with '0'
10
from blank to '_'
(ASCII)
87-900-986-01 (C)