INSTALLATION
!
The Turbo-V controller must be powered with 3-wire
power cord (see orderable parts table) and plug
(internationally approved) for user's safety. Use this
power cord and plug in conjunction with a properly
grounded power socket to avoid electrical shock and to
satisfy CE requirements. High voltage developed in the
controller can cause severe injury or death. Before
servicing the unit, disconnect the input power cable.
The Turbo-V controller can be used as a bench unit or
a rack module, but it must be positioned so that free air
can flow through the holes. Do not install or use the
controller in an environment exposed to atmospheric
agents (rain, snow, ice), dust, aggressive gases, or in
explosive environments or those with a high fire risk.
During
operation,
the
conditions must be respected:
•
temperature: from 5 °C to +45 °C
•
relative humidity: 0 - 95%
(without condensation)
See the appendix "Technical Information" for detailed
information about the above mentioned and the other
connections, and about the options installation.
USE
This paragraph describes the fundamental operating
procedures.
Detailed
procedures that involve optional connections or options
are supplied in the paragraph "USE" of the appendix
"Technical Information".
Make all vacuum manifold and electrical connections
and refer to Turbo-V pump instruction manual prior to
operating the Turbo-V controller.
WARNING!
NOTE
following
environmental
information
and
operating
To avoid injury to personnel and damage to the
equipment, if the pump is lying on a table make sure it
is steady. Never operate the Turbo-V pump if the pump
inlet is not connected to the system or blanked off.
The input signal J1 connector should be left in position
including the shipping links if no external connections
are made. The forepump and the Turbo-V pump can be
switched on at the same time.
Controls, Indicators and Connectors
The following paragraph illustrates the Controller front
panel and rear panel. More details are contained in the
appendix "Technical Information".
Front Panel Description
5
LS
AG
1
COUNTERS
increase
2
decrease
LOW SPEED
1. Keyboard push-button to recall on the display the cycle
number, cycle time and pump life.
2. Keyboard push-button for LOW SPEED mode selection.
It is active only when the front panel operation has been
selected. Pressed once, the pump runs at "stand-by"
speed. To unselect the mode, press the push-button
again.
3. Keyboard push-button to recall on the display the pump
current, pump temperature, pump power and rotational
speed. It is always active regardless of the operating
mode selected. Push-buttons 3 and 1, if pressed
together for at least 2 seconds, put the controller in a
routine where it is possible to program some operation
parameters.
4. Keyboard push-button for START, STOP/RESET mode
selection. It is active only when the front panel operation
has been selected. By pressing once the starting phase
begins; if pressed again it stops the pump. If the pump
has been stopped automatically by a fault, this push-
button must be pressed once to reset the controller and a
second time to restart the pump.
5. LCD back-lighted alphanumeric display: dot matrix 2
lines x 16 characters.
Controller Turbo-V 81-AG Front Panel
66
INSTRUCTIONS FOR USE
!
WARNING!
NOTE
R1
R2
VACUUM TECHNOLOGIES
MEASURES
next
menu
previous
START/STOP
Turbo-V 81-AG
87-900-986-01 (C)
3
4