Press the Print button to go into UCAL 1.
Using the Function key, scroll back to Position 26
.
Use the Print button in the input field to go into the decimal places.
Set the number of decimal places with the Tare button and
Function key.
Store with the Print button, the display goes to Position 27.
Use the Print button to go into the input field for the neutral
measurement factor.
Using the Tare button and Function key, set a number value in the
flashing space.
You can use the Zero button to move by one decade from right to
left.
After setting the "Neutral Measurement" factor, store with the Print
button, Position 01 is displayed.
Press the Zero button, appears on the display.
To save the setting and exit, press and hold the Print button then
briefly press the Zero button. The display returns to weighing
mode.
Neutral measurement application
Activate the "Neutral Measurement" function with the Function
key.
The display shows for "Neutral Measurement", confirm
with the Print button.
The display value appears along with the selected factor without a
unit weight. Stabilisation of the scales is represented on the
display by a triangle at the top left.
Press the ON/Off button to return to weighing mode.