Laboratorium

Advertenties

Beschikbare talen

Beschikbare talen

Display
• Excess voltage protection error
message
Display
• Excess current error message.
message
Display
• Overheat error message.
message
Display
• Voltage limit warning message
message
• The voltage setpoint cannot be
Display
• Current limit warning message
message
• The current setpoint cannot be
Any repair work other than that described above must be carried out by an authorised
technician. If you have questions about using the device, please contact our technical
support team.
message.
Voltage at the output is greater
than the set voltage setpoint.
The current at the output is
greater than the set current
setpoint.
System temperature is too high.
increased further because the
set voltage limit value has been
reached.
increased further because the
set current limit value has been
reached.
1. Switch the device off.
2. Disconnect consumers.
3. Switch the device back on.
4. If the error persists, contact
customer service.
1. Switch the device off.
2. Disconnect consumers.
3. Switch the device back on.
4. If the error persists, contact
customer service.
1. Switch the device off.
2. Disconnect consumers.
3. Allow the device to cool down for
30 minutes.
4. Switch the device back on.
5. If the error persists, contact
customer service.
• If necessary, increase the the
voltage limit value
• If necessary, increase the the
current limit value
49