Download Inhoudsopgave Inhoud Print deze pagina

Idmed WiTOF Gebruikershandleiding pagina 24

Inhoudsopgave

Advertenties

EMC Emission
Emission test
RF Emissions CISPR 11
RF Emissions CISPR 11
Harmonics IEC 61000-3-2
Voltage fluctuations
and flicker IEC 61000-3-3
EMC Immunity
Basic EMC
Phenomenon
standard
ELECTROSTATIC
D I S C H A R G E
61000-4-2
(ESD)
Radiated RF EM
Fields
61000-4-3
Proximity fields
from RF
wireless
com-
61000-4-3
munications
equipment
Electrical
fast
transients /
61000-4-4
bursts
Surges
Line-to-line
61000-4-5
Surges
Line-to-ground
61000-4-5
Gebruikershandleiding WiTOF
Compliance
Group 1
Class B
Class A
Compliant
Professional
healthcare
facility
environment
Immunity Test
Levels
± 8 kV contact
IEC
± 2 kV, ± 4 kV,
± 8 kV,
± 15 kV air
3 V/m
80 MHz – 2,7
IEC
GHz
80 % AM at
1 kHz
Complies to
IEC
table 9 of IEC
60601-1-2
(2014)
± 2 kV
IEC
100 kHz
repetition
frequency
IEC
± 0,5 kV, ± 1 kV
± 0,5 kV, ±
IEC
1 kV,
± 2 kV
EMC Instructions/cautions
The WiTOF uses RF energy only for internal functions.
Therefore RF emissions are very low and should not
disturb other nearby devices.
The WiTOF must be use in professional healthcare
facility environment
The WiTOF can be connected to the public mains
network
Compliance
levels
± 8 kV contact
In order to reduce ESD, the device
must be used in a 35% humidity en-
± 2 kV, ± 4 kV,
vironment or more
± 8 kV air
Use of this equipment adjacent to or
3 V/m
stacked with other equipment should
80 MHz – 2,7
be avoided because it could result in
GHz
improper operation. If such use is ne-
80 % AM at
cessary, this equipment and the other
1 kHz
equipment should be observed to ve-
rify that they are operating normally
Complies to
In order to prevent electromagnetic
table 9 of IEC
disturbance, keep minimum separati-
60601-1-2
on from RF communication equipment
(2014)
of 30cm
The WiTOF may temporarily not dis-
± 2 kV
play result during transient electro-
100 kHz
magnetic disturbances such as the
repetition
use of electrosurgery device. In that
frequency
case, the WiTOF will maintain the sa-
fety of the patient and the user.
Mains power quality should be that of
± 0,5 kV, ± 1 kV
a typical residential, commercial or
hospital environment.
Mains power quality should be that of
± 0,5 kV, ± 1
a typical residential, commercial or
kV, ± 2 kV
hospital environment.
EMC
Instructions/precautions
24

Advertenties

Inhoudsopgave
loading

Gerelateerde Producten voor Idmed WiTOF

Inhoudsopgave