Download Inhoudsopgave Inhoud Print deze pagina

Richtlijnen En Verklaring Van De Fabrikant - .bock belluno Montage- En Gebruikshandleiding

Inhoudsopgave

Advertenties

7

Richtlijnen en Verklaring van de fabrikant

Guidance and manufacturer's declaration
– Electromagnetic emission
The medizinisches Bett is intended for use in the electromagnetic environment specified below.
The customer or the user of the medizinisches Bett should assure that it is used in such an environment.
Emission test
RF emissions
CISPR 11 (partly)
RF emissions
CISPR 11 (partly)
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/flicker
emissions
IEC 61000-3-3
– Electromagnetic immunity
The medizinisches Bett is intended for use in the electromagnetic environment specified below.
The customer or the user of the medizinisches Bett should assure that it is used in such an environment.
Immunity test
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrostatic transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
Note: U
is the a. c. mains voltage prior to application of the test level.
T
Complliance
Electromagnetic environment - guidance
Group 1
The medical used bed uses RF energy only for its internal function. Therefore, its RF emissions are
very lowand are not likely to cause any interference in nearby electronic equipment.
Class B
Class A
The medizinisches Bett is suitable for use in all establishments other than domestic and those
directly connected to the public-voltage power supply network that supplies buildings used for
domestic purpose.
Complies
IEC 60601 test level
± 6 kV contact
± 8 kV air
± 2 kV for power supply lines
± 1 kV for input/output lines
± 1 kV differential mode
± 2 kV common mode
< 5 % U
T
(>95 % dip in U
) for 0.5 cycle
T
40 % U
T
(60 % dip in U
) for 5 cycles
T
70 % U
T
(30 % dip in U
) for 25 cycles
T
< 5 % U
T
(>95 % dip in U
) for 5 sec
T
3 A/m
Compliance level
± 6 kV contact
± 8 kV air
± 2 kV for power supply lines
± 1 kV for input/output lines
± 1 kV differential mode
± 2 kV common mode
< 5 % UT
(>95 % dip in UT ) for 0.5 cycle
40 % UT
(60 % dip in UT ) for 5 cycles
70 % UT
(30 % dip in UT ) for 25 cycles
< 5 % UT
(>95 % dip in UT ) for 5 sec
3 A/m
58
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the
relative humidity should be at least 30%.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment. If the user of
the medizinisches Bett requires continued
operation during power mains interruptions, it is
recommended that the medizinisches Bett be
powered from an uninterruptible power supply or
a battery.
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.

Advertenties

Inhoudsopgave
loading

Inhoudsopgave