Deviations from the specifications dictated in the PQ, are to be reported in the deviation report.
The PQ is concluded if all criteria of acceptance are approved and the possible deviations are rectified or
accepted.
Measurement – Prerequisites
ID
Description
P-1
The cabinet must be empty while conducting tests, ie without interior fittings
such as drawers, shelves etc.
Attachment:
Notes:
P-2
The measurements must be conducted in accordance to IEC 60068-3-5.
Attachment:
Notes:
P-3
The positioning of the sensors in the cabinet must be documented with a
sketch and/or a photograph.
Attachment:
Notes:
Name:
Conducted by:
Inspected / verified by:
74
Signature:
Model: ________________________
Accepted
Yes
Approved
Date:
(Yes / No):
SN: ________________________
No