The screen displays:
"Unit Overheat"
"Please Wait"
"Switch Off to"
"reset"
"menu"
The screen displays:
"Over Temp"
The screen displays:
"Heater Over Temp"
The screen displays:
"Unit Over Temp"
If your problem persists, it may be possible to rectify any issues over the
telephone. Our dedicated service and product support personnel may be
able to troubleshoot your problem remotely, thus causing minimal
disruption.
Contact our After Sales Department:
Please have to hand your model and serial number together with
information on the problem prior to contacting us.
If we are unable to solve your problem over the phone, we may suggest
returning your product to the distributor where purchased or directly to
Ultrawave. Where appropriate, we operate a Return To Base (RTB) warranty
and repairs policy.
Q-Series Operating Manual
The Q-Series has detected that the ultrasonic
generator on the unit has operated above the
rated temperature and has disabled the
ultrasonic function. This also disables a cycle
from starting. The most likely cause is the liquid
evaporating when the level sensor is switched
off. Switch the unit off, allow the unit to cool
for a few minutes, top up the liquid above the
level sensor and switch the unit back on.
The Q-Series has detected that the water in the
bath is above 75ºC and has aborted the cycle.
Please allow the liquid to cool or drain some
fluid and replace with cooler liquid.
The Q-Series has detected that the internal
water
heater
maximum temperature. Please allow the liquid
to cool or drain some fluid and replace with
cooler liquid.
The Q-Series has overheated due to prolonged
use. Please allow a few minutes for the unit to
cool.