Download Inhoudsopgave Inhoud Print deze pagina

Check Of Tightness; Connection And Use Of Outlet; List Of Medical Devices Compatible With Mediconnect-Duo- Line Connections - GCE MEDICONNECT-DUOLINE Gebruiksaanwijzing

Inhoudsopgave

Advertenties

6. OPERATION
6.1. BEFORE USE
VISUAL CHECK PRIOR TO THE INSTALLATION
Check the product for obvious damage (including labels and signs). If
there are marks of external damage, do not use the product and report
its status.
Visually check if product is clean. In case of pollution, clean it following
the procedure described in this instruction for use.
Using the marking system of GCE or the owner, check if lifetime or date
for disposal of the product has not been exceeded. If the date has been
exceeded, withdraw the product from the operation and report its status.
MEDICONNECT-DUOLINE is compatible for use with the gas stated on its
label ONLY! Never use it with di erent type of gas.

CHECK OF TIGHTNESS

After connection to the medical gas, check the product for visible or
audible leakage.
If you detect any leakage, use the below procedure „After use" and
return MEDICONNECT-DUOLINE for maintenance and repair.

6.2. CONNECTION AND USE OF OUTLET

LIST OF MEDICAL DEVICES COMPATIBLE WITH MEDICONNECT-DUO-
LINE CONNECTIONS
Hoses, flow meters, ventilators, low pressure regulators.
For information about the type of quick coupler for your product see AP-
PENDIX 1.
For Information about connecting / disconnecting procedure for your prod-
uct see APPENDIX 2.
Before connecting of any medical device, make sure your connector is
compatible with MEDICONNECT-DUOLINE outlet (see APPENDIX 1).
When more MEDICONNECT-DUOLINE outlets are used in the same time
may be the capacity of gas supply and performance of connected device
negatively a ected.
Make sure that the capacity of the equipment to be supplied by the
MEDICONNECT-DUOLINE is corresponding with capacity of the gas
source equipment. Take into account the pressure drop of the flexible
hoses.
6.3. AFTER USE
Disconnect the used medical device from MEDICONNECT-DUOLINE
and then disconnect MEDICONNECT-DUOLINE from gas supply.
5/96