Precision Novi™ Implantable Pulse Generator
11. Wipe the proximal end(s) clean.
CAUTION: Do not tunnel the splitter.
Note: If using the 2x8 Splitter and performing a permanent trial, the splitter tails may be
tunneled to the exit site.
Note: The following Codman Disposable Catheter Passers may be used in place of the
Boston Scientific tunneling tool:
REF 82-1515 (36 cm); REF 82-1516 (55 cm); REF 82-1517 (65 cm)
Note: When using a Codman Disposable Catheter Passer, tunnel from the midline incision to
the IPG pocket using the standard technique.

Connecting the Lead, Extension, Splitter, or Connector to the IPG

IPG ports are labeled as follows:
Connecting to the IPG:
• Leads connect to IPG ports C or D.
• For the Infinion 16, connect the splitter tail with laser-etched bands (contacts 1-8 of the
Infinion 16 lead) to the left port C and the unmarked splitter tail (contacts 9-16 of the
Infinion 16 lead) to the right port D.
• For the Infinion CX Lead, connect the tail with the single marker band (contacts 1-8 of
the Infinion CX Lead) to the left port C and the tail with two marker bands (contacts 9-16
of the Infinion CX Lead) to the right port D.
• For the Artisan 2x8 surgical lead, connect the left side to the left port C. Connect the
right side (the laser-etched tail), contacts 9-16, to the right port D.
Precision Novi™ Implantable Pulse Generator Directions for Use