Download Inhoudsopgave Inhoud Print deze pagina

Tunneling The Lead Or Lead Extension - Boston Scientific Precision Novi SC-1140 Gebruiksaanwijzing

Implantable pulse generator kit; implantable pulse generator and freelink rc kit
Inhoudsopgave

Advertenties

Precision Novi™ Implantable Pulse Generator

Tunneling the Lead or Lead Extension

Note: If using an Infinion™ CX Lead(s), it is recommended to use the Long Tunneling Tool
(35 cm).
1. If not already assembled, attach the tunneling tool handle to the shaft by turning the locking
mechanism clockwise.
Tool Handle
Locking Mechanism
2. Mark the desired route of the tunnel.
3. Administer the appropriate local anesthetic along the tunneling path.
4. OPTIONAL. If necessary, bend the tool shaft to conform to the patient's body.
5. Make a small incision at the desired exit site.
Precision Novi™ Implantable Pulse Generator Directions for Use
90930910-07
6 of 239
Shaft