UK
Guidance and manufacturer' s declaration – electromagnetic immunity
BRIZZY is intended for use in the electromagnetic environment specified below. The customer or the
user of BRIZZY should assure that it is used in such an environment.
IMMUNITY test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast transient/
burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short in-
terruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
34 MU_BRIZZY_REV6 / 07-2019 ©
IEC 60601 test level
± 6 kV contact
± 8 kV air
± 2 kV for power supply
lines
± 1 kV for input/output
lines
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
<5 % U
(>95 % dip in
T
U
) for 0,5 cycle
T
40 % U
T
(60 % dip in U
) for 5
T
cycles
70 % U
(30 % dip in U
T
for 25 cycles
<5 % U
(>95 % dip in
T
U
) for 5 s
T
3 A/m
Compliance
level
± 6 kV contact
± 8 kV air
Not applicable
Not applicable
Not applicable
)
T
3 A/m
Electromagnetic
environment – guidances
Floors should be wood, con-
crete or ceramic tile. If floors
are covered with synthetic
material, the relative humidity
should be at least 30 %.
Mains power quality should
be that of a typical commer-
cial or hospital environment.
Mains power quality should
be that of a typical commer-
cial or hospital environment.
Mains power quality should
be that of a typical commer-
cial or hospital environment.
If the user of BRIZZY requires
continued operation during
power mains interruptions, it
is recommended that BRIZZY
and the host computer be
powered from an uninterrupti-
ble power supply or a battery.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commer-
cial or hospital environment.