6.
Cleaning and maintenance
6.3
Automatic spool cycle
The automatic spool function is designed to prevent clogging of the soap and disinfectant tubes and to
prevent the waterline from being infected by Legionella and other waterborne bacteria. The automatic
spool function will initiate after a minimum amount of time the soap or disinfectant has not been in use.
This has been set by default at 2 weeks. If the soap or disinfectant has not been in use for 2 weeks, the
system will perform the following steps on the liquid that was last used 2 weeks ago:
•
Step 1: The soap or disinfectant pump will operate for 30 seconds
•
Step 2: Cold water flows for 60 seconds
Note: It is possible to change the frequency of the auto spool function for more frequent water flushing,
however it is not possible to change the frequency of the auto spool function for soap and disinfectant
flushing.
6.4
Replacing the pouch / bottle
Once the system does not dispense soap, lotion or disinfectant anymore, the pouch / bottle needs to be
replaced. Refer to fig. G / H for the correct installation of the pouch / bottle.
6.5
Priming the tubes
After the installation of the pouches/bottles, the tubes need to be primed. When the procedure has been
completed, the tubes are clean and free of air bubbles.
•
Press the LED switch on the water box for 1 second. The LED starts blinking.
•
To prime the soap tube: Activate sector 4, then sector 6. Activate sector 5 (sector 3 or 6 for
CLASSIC Pure S) to stop.
•
To prime the disinfectant tube: Activate sector 5, then sector 3. Activate sector 4 to stop.
•
Press the LED switch on the water box for 1 second to exit priming mode.