c) Measurement tolerances
Accuracy: ± (% of reading + count)
The accuracy is valid for one year at:
•
Ambient temperature: +3 °C (± 5 °C), ≤75 %, RH non-condensing
•
Operating temperature*: 18 to 28 °C (±1 °C)
*A temperature coefficient applies outside of this temperature range: +0.1 x (specified accuracy)
/ °C.
The measurement can be impaired when the device is operated within a high-frequency
electromagnetic field.
d) Continuity ( ) and diode (
Range
Resolution
0.1 Ω
1 mV
Overload protection: 600 V
e) DC voltage measurement
Range
200.0 mV
2.000 V
20.00 V
200.0 V
600 V
•
Input impedance: ≥100 MΩ for mV range (short circuit allows ≤5 digits), approx 10 MΩ
for other ranges.
•
Input voltage: max. 600 V
54
Remark
•
Open circuit: Resistance >100 Ω, no beep.
•
Circuit with a good connection: Resistance ≤10 Ω, con-
secutive beeps.
•
Open circuit voltage: Approx. 2.2 V
•
Silicon PN junction voltage: Appox. 0.5 to 0.8 V
Resolution
0.1 mV
0.001 V
0.01 V
0.1 V
1 V
) test
Accuracy
± (0.5 % + 8)
± (0.8 % + 7)