In order to prevent the cables from being damaged when
activating the motors, follow the instructions below.
1. Connect the cables of the bed end motors (head and
foot end) directly to the control box.
2. The plug of the thigh motor must be directed through
the opening of the supporter for the backrest motor and
then connected to the control box.
3. Connect the main cable plug to the power socket.
4. Run the motors of the bed ends to their top position.
5. Place the motor cable of the foot end on the hook at
the foot end.
6. Run the backrest to its top position.
1551227-F
7. Place the head end motor cables on the four hooks at
the head end.
8. Attach the locking cam over the plugs in the control box.
3.5 Dismantle the bed
1. Dismantle side rails and lifting pole.
2. Bring the bed to its lowest position and adjust all
mattress support sections to horizontal position.
3. Disconnect the 230 V supply. Roll the cable onto the
hook of the head end of the bed.
4. Disconnect all the motor plugs from the motors (do not
disconnect the backrest motor cable).
5. Dismantle the locking cam from the control by means of
a tool, e.g. a flat headed screwdriver.
6. Pull out the cables of the bed end motors and leg
section motor.
7. Separate the mattress support from the bed ends.
8. Divide the head and foot end of the mattress support.
9. For beds with 24 V external power supply, verify that
mains cable is attached to control unit power supply
cable.
Setup
9