Etude Plus
Degree of protection: IPx6* or IPx4** (depending on
configuration)
The control unit, external power supply and motors are
protected according to IPx6. When the bed is equipped
with a HL80 hand control, it is protected according to IPx4
Insulation class: Class II
Type B Applied Part
Applied Part complying with the specified requirements for
protection against electrical shock according to IEC60601-1.
(An applied parts is a part of the medical equipment which
is designed to come into physical contact with the patient
or parts that are likely to be brought into contact with
the patient.)
Sound level: 45 to 50 dB (A)
* IPX6 classification means that the electrical system is
protected against water projected in powerful jets from any
direction.
20
** IPX4 classification means that the electrical system is
protected against water splashed against the component
from any direction.
The bed is not provided with a mains switch, so the
mains plug is the only separation from the mains.
1551227-F