Benaming
Max. measuring cell tem-
perature
Actual resonance fre-
quency
Min. resonance frequency
Max. resonance frequency 306
Probe
Output
Output 2
Device name
Serial number
Hardware version
Software version
Device revision
Simulation switching out-
put
Simulation value output
Simulation switching out-
put 2
Simulation value output
Device status detailed
status
Systeemcommando's
Benaming
Factory Reset
Reset Pointer - Resonance Frequency
Reset Pointer - Measuring Cell Temperature
Reset Pointer - Electronic Temperature
Uncovered
VEGAPOINT 23 • Transistor meit IO-Link
ISDU (dez)
ISDU (hex)
303
0x011E
304
0x0130
305
0x0131
0x0132
307
0x0133
308
0x0134
309
0x0135
310
0x0136
311
0x0137
312
0x0138
313
0x0139
314
0x013A
315
0x013B
316
0x013C
317
0x013D
318
0x013E
319
0x013F
Grootte
Datatype
(Byte)
4
Float
4
Float
4
Float
4
Float
2
Unsigned16 RO
2
Unsigned16 RO
2
Unsigned16 RO
19
String
16
String
19
String
19
String
2
Unsigned16 RO
1
Unsigned8
2
Unsigned16 RW
1
Unsigned8
2
Unsigned16 RW
4
Unsigned32 RO
ISDU (dez)
ISDU (hex)
130
0x082
161
0x0A1
163
0x0A3
164
0x0A4
165
0x0A5
13 Bijlage
Toegang Waardebereik
RO
-
RO
-
RO
-
RO
-
0 = Not Covered
256 = Covered
512 = Covered in-
side Window
768 = Covered
outside Window
0 = Open
1 = Closed
0 = Open
1 = Closed
RO
-
RO
-
RO
-
RO
-
-
RW
0 = Off
1= On
0 = Open
1= Closed
RW
0 = Off
1= On
0 = Open
1= Closed
0 = Open
1= Closed
Toegang
WO
WO
WO
WO
WO
51