1 Specifications
on page 3
2 General information
on page 4
3 What you have received
Section 1 Specifications
The product has only the approvals listed and the registrations, certificates and declarations officially
provided with the product. The usage of this product in an application for which it is not permitted is
not approved by the manufacturer.
Specifications are subject to change without notice.
Specification
Type
Dimensions (Ø × L)
Weight
Other
Certification
Note: The intrinsically safe C110E ATEX sensors have no smart capability to store calibration data. All C110E
ATEX sensors have the ATEX conformity information engraved on the sensor.
Note: ATEX sensors can only be used with ATEX certified portable instrument 3650EX and wall mount instruments
366xEX.
Specification
Recommended applications
Material
Thickness [μm]
Calibration gas
Dissolved measurement range
1
Accuracy
Expected current in air @ 1 bar 25°C
[μA]
Temperature compensation range
Temperature measuring range
2
Response time
1
At room temperature (25 °C).
2
Response time at 25°C for a 90% signal change
on page 5
Non-ATEX sensors
Smart capability
CE
Membrane 2956A
Trace measurement
0 ppb to 50 ppm
The greater of ±1% of reading
(± 5% for sensors calibrated in air)
4 Installation
on page 6
5 Maintenance
on page 8
6 Troubleshooting
on page 12
Electrochemical ozone sensor
39.5 × 86.2 mm (1.56 × 3.39 in.)
300 g
PFA
25
Span gas or air
or ± 0.4 ppb,
or ±1 Pa
25.3
– 5 to 45° C
– 5 to 100° C
25 sec.
ATEX sensors
Intrinsically safe
CE, Ex II 1 G, Ex ia IIC T6
Membrane 29552A
High concentration (> 1 mg/l)
PTFE
50
0 ppb to 200 ppm
The greater of ±1% of reading
(± 5% for sensors calibrated in air)
or ± 20 ppb,
or ± 4 Pa
6.5
6 min.
English 3