Setting The Output Voltage (Remote Voltage Programming); Maintenance - Simco ECM 30 Gebruikershandleiding

Advertenties

+
Attention:
When the remote on/off function is not (no longer) required:
-
Remove the connection between pins 24 and 25 on the Sub-D connector or remove
the Sub-D connector.

8.3. Setting the output voltage (remote voltage programming)

8.3.1. Summary
The output voltage set point can be set remotely (Remote Voltage Programming). This option is
connected using the Sub-D connector on the rear of the unit.
8.3.2. Installation
To activate the Remote Voltage Program function, pins 17 and 18 of the Sub-D connector must
first be connected together. Next, a power supply can be connected between pin 3 (+) and pin
16 (GND). The output voltage can now be externally regulated from 0 kV to maximum with 0 to
10 V.
(see also the diagrams in the appendix)
+
Attention:
-
When the Remote Voltage Program function is not (no longer) required:
Remove the connection between pins 17 and 18 on the Sub-D connector or remove
the Sub-D connector.

9. Maintenance

- Keep the unit dry and free of dust, dirt and chemicals.
- Regularly check the generator earthing.
- Regularly check the high-voltage connectors and cables for mechanical or electrical damage.
- The generator does not contain any parts that require periodic maintenance.