Cap is Closed, Locked and Watertight
The following picture show the proper steps to open and close the battery cap.
Steps to Open or Close the Cap
The following pictures shows the proper steps to utilizing the air pump.
1. Remove the Valve Plug
3. Remove the Pump Tube
*FFFT* sound is residual air escaping the pump hose. This is normal - no air is lost from the Knee.
•
Pump tip must be clean and free of debris and lint before inserting.
•
Knee must be fully extended when adjusting air pressure.
•
After re-attachment, gauge reads about half of the pressure in the knee.
C. Alignment and Setup
Follow the Prosthetist Setup Guide to properly align and adjust the device. Two installation programs, Plié Control and Plié
Gait Lab, are provided to optimize the computer settings for each patient. Verify the computer system minimum requirements
before software installation (Microsoft Windows 7 or later operating system, 1.0 MHz Processor, 256K RAM and USB Port).
D. Alignment
Careful attention to the alignment of the socket in relation to the Plié® 3 MPC Knee and the prosthetic foot is essential for a
successful user outcome. The prosthesis alignment should account for the range of motion (ROM), voluntary control, and
balance of individual users. Proper alignment and user voluntary control are essential to the optimal function of the prosthesis.
To Open: Press down latch (the cap will spring open).
To Close: Press firmly on entire cap until latch engages.
2. Insert Pump Tube and Pump Air
4. Re-insert Valve Plug
Cap is Unlocked and Opened