Information on standards
5. Product description
6. Commissioning
Inserting batteries
This product complies with the EN 13485 standard for
penetration measurement.
Suitability: S, T (storage, transportation)
Environment: E (transportable thermometer)
Accuracy class: 0.5
Measuring range: -50...+250 °C
According to EN 13485, the measuring instrument should
be checked and calibrated regularly under the terms of
EN 13486 (recommended frequency: yearly).
Contact us for more information.
1
Infrared sensor
2 2-point laser
3 Display
4 Control keys:
· [ON]: switches the instrument on
· [OFF]: switches the instrument off
(hold button down)
· [
]: switches to IR measurement,
carry out IR measurement (hold button
down)
· [
]: switches to contact measurement
· [HOLD/MIN/MAX]: holds measuring
value, displays minimum/maximum
value
5 Fold-out immersion/penetration probe,
folding out the probe switches the
instrument on
6 Battery compartment (at the back)
1 Use a slotted screwdriver to undo the
screw on the battery compartment.
2 Open the battery compartment.
3 Insert batteries (2x type AAA). Observe
the polarity!
4 Close the battery compartment.
5 Tighten the screw.