Guidance and manufacturer's declaration – electromagnetic immunity
The UPA-WU10T/UPA-WU10R is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. Portable RF communications equipment should be used
no closer than 30 cm (12 inches) to any part of the UPA-WU10T/UPA-WU10R. Otherwise,
degradation of the performance of this equipment could result.
Immunity test
Proximity fields
from RF wireless
communications
equipment
IEC 61000-4-3
NOTE:
These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a For some services, only the uplink frequencies are included.
10
GB
a
Band
Service
380 –
TETRA 400
390 MHz
430 –
GMRS 460
470 MHz
FRS 460
704 –
LTE Band 13, 17
787 MHz
GSM 800/900
TETRA 800
800 –
iDEN 820
960 MHz
CDMA 850
LTE Band 5
GSM 1800
CDMA 1900
GSM 1900
1,700 –
DECT
1,990 MHz
LTE Band 1, 3,
4, 25
UMTS
Bluetooth
WLAN
2,400 –
802. 11 b/g/n
2,570 MHz
RFID 2450
LTE Band 7
5,100 –
WLAN 802. 11
5,800 MHz
a/n
IEC 60601
a
Modulation
test level
Pulse
modulation
27 V/m
18 Hz
FM
±5 kHz
28 V/m
deviation
1 kHz sine
Pulse
modulation
217 Hz
Pulse
modulation
28 V/m
18 Hz
Pulse
modulation
28 V/m
217 Hz
Pulse
modulation
28 V/m
217 Hz
Pulse
modulation
217 Hz
Compliance
level
27 V/m
28 V/m
9 V/m
9 V/m
28 V/m
28 V/m
28 V/m
9 V/m
9 V/m