4.4.2 Performing a Calibration

Please prepare the T-CAL
calibration. Make sure that the T-CAL
blemishes.
• T-CAL
standards for TB 350 IR: 20 / 800 / 4000 NTU
®
• T-CAL
standards for TB 350 WL: 20 / 800 / 2000 / 4000 NTU
®
1. Zero Adjustment:
Sets zero standard.
2. Prepare Vial:
Oil sample vial if necessary.
3. Insert Standard:
Insert standard value specified on screen.
User will insert series of various turbidity standards to complete
calibration.
4. Modal:
Represents active process.
5. Log Summary:
Press Save Button to save log.
Press Cancel Button to return to beginning of calibration
process.
EN   TB 350
standards included in the scope of delivery before
®
vials are free of stains, scratches or
®
4 Operation
EN
17