4.3.2 Performing a Measurement
Please prepare the sample cells included in the scope of delivery before
measurement. Make sure that the sample cells are free of stains, scratches
or blemishes.
1. Collect sample
Ensure sample is representative (well mixed).
Glassware should be free of stains, scratches or other
blemishes.
Rinse collection vessel at least two times with sample
prior to sample collection.
2. Prepare sample cell
Using a lint free cloth, dry the exterior of the sample
vial.
Use silicone oil to minimize stray light reflections.
3. Insert vial into instrument
Ensure sample is bubble free.
Invert slowly to mix. Do not shake.
4. Close lid
Measurement will not be allowed without lid in place.
5. Check settings
Ensure the measurement settings are correct.
Settings for measurement cannot be changed after
measurement occurs.
6. Press measure button
Starts Measurement.
7. Data Log
Enter Data Log to save measurement.
EN TB 350
4 Operation
EN
15