6 Maintenance
6.1 General Maintenance Information
IMPORTANT !
Service and maintenance of the Soft Tilt must be
carried out together with the bed.
– For detailed information on maintenance
procedures and checklists see Bed Service
Manual for the bed in use.
6.2 Cleaning and disinfection
IMPORTANT!
Wrong fluids or methods can harm or damage
your product.
– Never use corrosive fluids (alkalines, acid,
cellulose thinner, acetone etc)
– Never use a solvent that changes the structure
of the plastic or dissolves the attached labels.
Electrical components
IMPORTANT!
The IP classification determines the washability of
the electrical components.
Electronics classified IPx5 and/or IPx6 may NOT
be washed with jet based cleaning equipment or
in a washing tunnel.
– The IP classification is stated on the serial labels
of the electrical components.
– Please also note that the components can have
different classification.
1605513-C
Textiles
See attached label on the textile cover for detailed washing
instructions
General Cleaning Method
Method: Wipe off with a wet cloth or soft brush.
Max temp: 40 °C
Solvent/chemicals: Mild household detergent or soap and
water.
IMPORTANT
For detailed information of cleaning methods
see instructions in the section Non-washable
components in the Bed User Manual
6.3 Battery
We recommend a check of the battery every 6 months:
1. Elevate both wings and unplug the power cable.
2. Press ON/OFF & CPR button A and ensure both wings
move into horizontal position.
We recommend replacement of the battery after 4 years –
Call dealer/technician to replace battery.
Storage
IMPORTANT
Before storage of the Soft Tilt, ensure the
system has been turned off completely and the
emergency stop button is pressed to prevent the
battery from discharging.
Maintenance