Max./min. values are not recorded in a data set.
Only the current max./min. values are displayed in the PC software.
ALARM SETTING MODE:
Press and hold the ALM button for the three seconds.
HIGH AL RH for the selected channel appears in the display.
The upper alarm limit for the humidity is flashing.
With the ALM button you can select one possible alarm limit
after the other:
- High limit humidity („HIGH AL RH", default 70%)
- Low limit humidity („LOW AL RH", default 20%)
- High limit temperature ("HIGH AL °C", default 40°C)
- Low limit temperature („LOW AL °C", default 0°C)
When displayed, the limit value can be changed by pressing
the "+" or "-" button.
In order to activate an alarm limit, the "OK/MAX.MIN." button
must be pressed while the corresponding limit value is
displayed.
If the alarm limit is activated, the alarm on/off symbol is
displayed as active on the toolbar (not crossed out).
If the alarm limit is not activated, the alarm on/off symbol is
displayed as inactive on the toolbar (crossed out).
Press and hold + or - button in setting mode, you will enter
fast mode.
To enter normal mode again, press either ALM button again
or press for at least 20 seconds no button.
Alarm function:
In case of an alarm, Symbol ALARM (above the date) and
the corresponding channel number flashes. Also the Hi or Lo
Symbol is flashing if the corresponding channel is shown.
If the alarm sound is activated, the symbol (bell) flashes and
the acoustic alarm will sound for 2 minutes.
Press ALM button to stop the alarm sound.
The corresponding "Hi" or "Lo" Symbol and ALARM continue
to flash until the corresponding value is again within the
alarm limit or the alarm limit is deactivated in alarm setting
mode.
Note:
42