Electromagnetic Compatibility (EMC, IEC 60601-1-2:2007, Table 2)
Electromagnetic immunity
The electric breastpump Swing Maxi is intended for use in the electromagnetic environment
specified below. The customer or the user of the electric breastpump Swing Maxi should
assure that it is used in such an environment.
Immunity test
Electrostatic
discharge ( ESD )
IEC 61000-4-2
Electrical fast
transient / burst
IEC 61000-4-4
Surge
IEC 61000-4-5
All manuals and user guides at all-guides.com
IEC 60601
Compliance
test level
level
± 6 kV contact
± 6 kV contact
± 8 kV air
± 8 kV air
± 2 kV for power
± 2 kV for power
supply lines
supply lines
± 1 kV for input /
± 1 kV for input/
output lines
output lines
± 1 kV line( s ) to
± 1 kV differential
line(s)
mode
± 2 kV line( s ) to
± 2 kV common
earth
mode
Electromagnetic
environment
– guidance
Floors should be wood,
concrete or ceramic
tile. If floors are covered
with synthetic material,
the relative humidity
should be at least 30%.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment.