4.4. On / Off switch
On the right hand side of the cover cap, there is the on /
off switch (See figure 16) . If the chair is out of use for an
extended period, you are advised to turn it off . This will
prevent any unauthorised use of the chair .
5. Cleaning and maintenance
Figure 16 On / off switch .
The Elexo XXL must be cleaned after every use, with water and regular household
cleaning agents . Remove excess soap residues . PUR integrated foam components with
IMC coating should be rubbed clean and disinfected . Do not use any abrasive or other
aggressive cleaning products (e .g . chlorine bleach) . Practical experience has shown that
regular household cleaning agents with limited alcohol content (max . 5%) are most
suitable for this purpose . The product Neoquat combi
®
is recommended for disinfection
purposes (on the basis of long-term testing) . If necessary, the PUR components can be
cleaned with a solution of 70% ethanol or propanol .
You must ensure that both the disinfecting and cleaning processes involve the
components being rubbed and not just submerged . The time the products must be left
on the components has a significant impact upon the lifespan of the PUR components .
As a result of the material properties of PUR, contact with sharp objects must be
avoided . This could lead to damage or tears in the material . Finally, we would like
to point out that non-critical parts of the PUR components must also be cleaned and
decontaminated .
Check the wheels regularly for hairs, dust and soap residues and remove when
necessary . Maintenance and repairs to the Elexo XXL may only be carried out by
authorised personnel .
Elexo XXL Shower-toilet chair | Instructions for use | 35